Abstract:

Considering the wear in mechanical bearings and the requirement for sensors or electrical power in active magnetic bearings, a novel nutation blood pump using a passive magnetic spherical bearing was developed in this paper. A mathematical model was derived to calculate the magnetic forces of the bearing between two pairs of magnetic sleeves in the nutation process. The calculation results demonstrate that the fluctuations of magnetic forces enlarge with the increase in the nutation angle; the magnetic forces obviously increase with the decrease in the air gap, especially along the z-axis. The dynamic magnetic finite element simulation was carried out to validate the mathematical model. The simulation and calculation results of the magnetic forces show consistent trends and provide a theoretical basis for the parameter design. To validate the performance of the pump, computational fluid dynamics (CFD) analyses and in vitro experiments were conducted. The predicted values of the flow velocity vector through the pump, and the wall shear stress, demonstrate that the pump has an antithrombotic property and would not cause serious blood damage. The hydraulic experiment shows that a pressure rise can be achieved in the range of 60140 mmHg, at a rotational speed of 6001600 rpm and a flow rate of 0.46.7 L/min. The normalized index of hemolysis (NIH) of the nutation pump was 0.0043 0.0008 g/100L. The in vitro tests indicate the feasibility of a magnetically levitated ventricular-assist nutation blood pump for further suspension stability and animal trials.

Keyword:

bearing capacity Blood Heart Magnetic confinement Magnetic forces Magnetic levitation numerical simulation Nutation blood pump passive magnetic levitation Permanent magnets Pumps ventricular assist device

Community:

  • [ 1 ] [Chen, Gang]Fuzhou Univ, Sch Mech Engn & Automat, Fuzhou 350116, Peoples R China
  • [ 2 ] [Yao, Ligang]Fuzhou Univ, Sch Mech Engn & Automat, Fuzhou 350116, Peoples R China
  • [ 3 ] [Zheng, Rongye]Fuzhou Univ, Sch Mech Engn & Automat, Fuzhou 350116, Peoples R China
  • [ 4 ] [Zhang, Long]Fuzhou Univ, Sch Mech Engn & Automat, Fuzhou 350116, Peoples R China
  • [ 5 ] [Ding, Jiaxin]Fuzhou Univ, Sch Mech Engn & Automat, Fuzhou 350116, Peoples R China
  • [ 6 ] [Chen, Gang]Sanming Univ, Sch Mech & Elect Engn, Sanming 365004, Peoples R China

Reprint 's Address:

  • 姚立纲

    [Yao, Ligang]Fuzhou Univ, Sch Mech Engn & Automat, Fuzhou 350116, Peoples R China

Show more details

Related Keywords:

Source :

IEEE ACCESS

ISSN: 2169-3536

Year: 2019

Volume: 7

Page: 169327-169337

3 . 7 4 5

JCR@2019

3 . 4 0 0

JCR@2023

ESI Discipline: ENGINEERING;

ESI HC Threshold:150

JCR Journal Grade:1

CAS Journal Grade:2

Cited Count:

WoS CC Cited Count: 4

SCOPUS Cited Count: 4

ESI Highly Cited Papers on the List: 0 Unfold All

WanFang Cited Count:

Chinese Cited Count:

30 Days PV: 2

Online/Total:213/10036233
Address:FZU Library(No.2 Xuyuan Road, Fuzhou, Fujian, PRC Post Code:350116) Contact Us:0591-22865326
Copyright:FZU Library Technical Support:Beijing Aegean Software Co., Ltd. 闽ICP备05005463号-1