Indexed by:
Abstract:
A novel method for the determination of Cr(III) and Cr(VI) in gelatin capsules is reported that involved separation by capillary electrophoresis and the determination using inductively coupled plasma mass spectrometry with a collision-reaction cell. The separation was achieved on a fused-silica capillary column (75 μm ID × 30 cm length). The electrophoretic buffer was 20 mM NaH2PO4-5 mM Na2B4O7 (pH 6.0) with an applied voltage of 15 kV. Chromium species in the capsules were extracted into a buffer-5 mM 1,2-cyclohexanediaminetetraacetic acid mixture. The spike recoveries of chromium species were in the range of 87% to 95%. The detection limits of Cr(III) and Cr(VI) were 0.61 and 0.45 ng · mL-1, and the procedure was applied for the speciation of capsules. Approximately 0.242-71.64 μg g-1 of Cr(III) were detected in the capsules. © 2014 Taylor & Francis Group, LLC.
Keyword:
Reprint 's Address:
Email:
Source :
Analytical Letters
ISSN: 0003-2719
Year: 2014
Issue: 14
Volume: 47
Page: 2406-2416
1 . 0 3
JCR@2014
1 . 6 0 0
JCR@2023
ESI HC Threshold:268
JCR Journal Grade:4
CAS Journal Grade:4
Cited Count:
SCOPUS Cited Count: 5
ESI Highly Cited Papers on the List: 0 Unfold All
WanFang Cited Count:
Chinese Cited Count:
30 Days PV: 2
Affiliated Colleges: