Abstract:
Hot electron induced cathodic electrochemiluminescence (ECL) was found at a disposable CdS modified screen printed carbon electrodes (CdS-SPCEs) during cathodic pulse polarization. The ECL behavior of the modified electrode was investigated with two emissions located at 520 nm and 580 nm. The CdS-SPCEs show very stable and reproducible ECL signal. The ECL intensity was linear with the dissolved oxygen concentration in the range of 1.7-33 mg/L with a detection limit of 0.02 mg/L (S/N = 3). In addition, a possible mechanism for ECL behavior of the modified electrode was proposed. The developed method can be applied to detect the dissolved oxygen concentration or biochemical oxygen demand (BOD). © 2011 Elsevier B.V.
Keyword:
Reprint 's Address:
Email:
Source :
Sensors and Actuators, B: Chemical
ISSN: 0925-4005
Year: 2011
Issue: 2
Volume: 157
Page: 488-493