Indexed by:
Abstract:
A novel form of reversed-phase liquid chromatography (RPLC) by the dynamically modified hydrophilic interaction monolithic column has been described in this paper. A porous poly(SPMA-co-PETA) monolith with strong cation-exchange (SCX) was prepared and the resulting monolith showed a typical hydrophilic interaction chromatography (HILIC) mechanism at higher organic solvent content (ACN% > 50%). The good selectivity for neutral, basic and acidic polar analytes was observed in the HILIC mode. In order to increase the hydrophobic interaction, the monolith with SCX was dynamically modified with a long-chain quaternary ammonium salt, cetyltrimethylammonium bromide (CTAB), which was added to the mobile phase. CTAB ions were adsorbed onto the surface of the SCX monolithic material, and the resulting hydrophobic layer was used as the stationary phase. Using the dynamically modified SCX monolithic column, neutral, basic and acidic hydrophobic analytes were well separated with the RPLC mode. Crown Copyright (C) 2009 Published by Elsevier B.V. All rights reserved.